Clinical experience with the Fieldmaster perimeter
American Journal of Ophthalmology
|February 1, 1982
Summary
Fieldmaster testing accurately detects 95% of visual field defects compared to Goldmann perimetry. This method is reliable for clinical practice, even with less experienced staff, showing high diagnostic value.
Area of Science:
- Ophthalmology
- Visual Science
Background:
- Visual field testing is crucial for diagnosing and monitoring various eye conditions.
- Goldmann kinetic perimetry is a standard but time-consuming method.
Purpose of the Study:
- To evaluate the efficacy and reliability of the Fieldmaster Model 200 for visual field testing in a clinical setting.
- To compare Fieldmaster testing with Goldmann kinetic perimetry.
Main Methods:
- 167 consecutive patients underwent perimetry using the Fieldmaster Model 200.
- Results were compared against Goldmann kinetic perimetry findings.
Main Results:
- Fieldmaster testing identified 95% of defects found by Goldmann perimetry.
- False-positive and false-negative rates were minimal.
- Unskilled personnel could perform the test effectively.
- Incomplete examination rates decreased to <2% with familiarity.
Conclusions:
- Fieldmaster Model 200 is a highly effective and reliable tool for visual field testing in clinical practice.
- It offers comparable diagnostic accuracy to Goldmann perimetry with improved efficiency.
- The test is suitable for use by various personnel, enhancing accessibility.