Warfarin anticoagulation: difficulties in interpretation of the prothrombin time
Thrombosis Research
|January 1, 1982
Summary
Blood collection tube material significantly affects prothrombin time (PT) in patients on warfarin. Non-polypropylene tubes cause falsely shortened PT, risking incorrect warfarin derivative dosage.
Area of Science:
- Clinical Chemistry
- Hematology
- Pharmacology
Background:
- Prothrombin time (PT) is a critical measure for monitoring warfarin therapy.
- Accurate PT measurements are essential for safe and effective anticoagulation.
- In vitro factors can influence coagulation test results, potentially impacting clinical decisions.
Purpose of the Study:
- To investigate the effect of blood collection tube material on prothrombin time measurements.
- To determine if collection tube type influences PT in patients receiving warfarin or coumadin therapy.
- To assess the clinical implications of altered PT results due to collection tube variations.
Main Methods:
- Prothrombin time was measured in normal individuals and patients on warfarin therapy.
- Blood samples were collected using various types of blood collection tubes.
- The influence of time and temperature on PT shortening was evaluated for coumadin-treated patients.
Main Results:
- A significant shortening of prothrombin time was observed when blood was collected in non-polypropylene tubes.
- This effect was noted in both normal subjects and patients undergoing warfarin therapy.
- The in vitro shortening of PT was dependent on both collection time and temperature.
Conclusions:
- The material of the blood collection tube is a critical pre-analytical variable affecting PT results.
- Using non-polypropylene collection tubes can lead to falsely low PT values in patients on warfarin.
- These findings highlight the potential for serious clinical errors in warfarin derivative dosage due to improper blood collection.