Related Experiment Video
Updated: Aug 13, 2026

Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart
Published on: May 11, 2018
A simple left heart assist device for use after intracardiac surgery: development, deployment and clinical experience
Abstract:
A simple left heart assist device (LHAD) has been developed and employed in nineteen patients with severe left ventricular dysfunction who could not be weaned from cardiopulmonary bypass following intracardiac surgery. It has been used when all other means of weaning, including maximum pharmacologic therapy and intra-aortic balloon counterpulsation (IABC), had failed. The device utilizes specially designed and constructed obturated cannulae in the left atrium and the ascending aorta, and an extracorporeal roller pump to partially bypass the left ventricle. With improved cardiac performance, the patient may be separated from the device without need for thoracic reentry. Of the nineteen patients having LHAD support (2-500 hours), thirteen were eventually weaned from the device and seven were discharged from the hospital. Five patients remain alive and well (18 to 50 months postoperatively).
More Related Videos
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
05:18Implantation of Left Ventricular Assist Device (LVAD) in Juvenile Landrace Swine: A LVAD Implantation Model of Pediatric Heart Failure
Published on: January 16, 2026
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiomyopathy V: Interprofessional Care