Related Experiment Videos
Systematic error associated with apparatus 2 of the USP dissolution test II: Effects of deviations in vessel
Journal of Pharmaceutical Sciences
|April 1, 1982
Abstract:
Dissolution vessels made from glass or plastic are recognized by the USP as being suitable for dissolution testing. Glass vessels with a bottom inside curvature flatter than that of a sphere can cause a high bias in dissolution results; vessels with a steeper curvature can cause a low bias. The inside bottom curvature of plastic vessels adhered closely to the curvature of a sphere. The plastic vessels are preferable for use if the drug is not adsorbed and the vessel is not attacked by the dissolution medium. Bias in results between individual positions of a dissolution apparatus was traced to two shafts which were not vertical.