Related Experiment Video
Updated: Oct 24, 2026

Automated Modular High Throughput Exopolysaccharide Screening Platform Coupled with Highly Sensitive Carbohydrate Fingerprint Analysis
Published on: April 11, 2016
A chromatographic quality control procedure based on HPLC for 5-hydroxymethylfurfural in autoclaved D-glucose
Abstract:
A specific qualitative chromatographic assay method is described for 5-hydroxymethylfurfuraldehyde (5HMF) in degraded D-glucose solutions incorporating an internal standard. Comparison is made between the chromatographic method and the official direct ultraviolet absorption measurements at 284 nm for solutions of D-glucose degraded to different extents. Hospital samples of autoclaved Dextrose Injection, 5% w/v, are also examined by both methods and it is demonstrated that quite different results can be obtained depending upon the conditions of measurement. It is demonstrated that the direct absorption method does not provide an adequate limit test for compounds related to 5HMF.
Related Concept Videos
High-Performance Liquid Chromatography: Instrumentation
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...