Automated sampling of in vitro dissolution medium: effect of sampling probes on dissolution rate of prednisone
Abstract:
The effect of sampling probe size and location on the in vitro dissolution rate of prednisone tablets was examined. Using USP XX Apparatus 2 with an automated sampling system, dissolution rates were determined using two types of large filter-tipped probes and a small capillary probe. Each probe was tested at three locations within the kettle. The large probes caused hydrodynamic changes which, when compared with results obtained through manual sampling, resulted in significant changes in dissolution rates at each location. These changes were less evident when the capillary probe was used, with an insignificant difference between results of automated and manual sampling when the capillary probe was placed midway between the paddle shaft and the kettle wall.