The feasibility of performing 133Xe ventilation imaging following the perfusion study
Radiology
|August 1, 1982
Abstract:
The 99mTC Compton scatter contribution from the perfusion lung study to the subsequent 133Xe ventilation scan was quantitated in 15 consecutive patients undergoing both studies. Maximum contribution during the wash-in phase occurred during the first 10 seconds; thereafter, 99mTc Compton scatter averaged less than 10% of the total counts accumulated for a 133Xe image. The contribution during wash-out gradually increased to a maximum of 36% at 4 minutes in a normal subject. The authors conclude that a properly performed 133Xe ventilation study may immediately follow the 99mTc perfusion lung scan without significant image degradation.