Titanium implants permanently penetrating human skin

P I Brånemark, T Albrektsson

    Scandinavian Journal of Plastic and Reconstructive Surgery
    |January 1, 1982
    PubMed
    Summary

    Non-alloyed titanium implants, when designed to minimize skin movement, show no adverse effects for permanent transcutaneous passage. This biocompatible material and surgical technique offer promising solutions for skeletal-anchored prostheses.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Two-piece Zirconia Implants: An Office based Retrospective Study with up to 7 and mean 3 Year Follow-Up.

    The International journal of oral & maxillofacial implants·2025
    Same author

    Osteoimmune regulation underlies oral implant osseointegration and its perturbation.

    Frontiers in immunology·2023
    Same author

    Is the intake of selective serotonin reuptake inhibitors associated with an increased risk of dental implant failure?

    International journal of oral and maxillofacial surgery·2017
    Same author

    Response to the Letter to the Editor: "Factors Influencing Early Dental Implant Failures"

    Journal of dental research·2016
    Same author

    Bruxism and dental implant failures: a multilevel mixed effects parametric survival analysis approach.

    Journal of oral rehabilitation·2016
    Same author

    Turned versus anodised dental implants: a meta-analysis.

    Journal of oral rehabilitation·2016

    Area of Science:

    • Biomaterials Science
    • Medical Engineering
    • Surgical Innovation

    Background:

    • Transcutaneous implants require careful consideration of material, design, and surgical technique to ensure long-term biocompatibility and function.
    • Previous attempts at skin-penetrating implants have faced challenges with soft tissue integration and inflammatory responses.

    Purpose of the Study:

    • To investigate the conditions necessary for successful permanent transcutaneous passage of titanium implants.
    • To evaluate the soft tissue response and long-term stability of titanium implants used for skeletal anchorage.

    Main Methods:

    • Clinical investigation involving 8 volunteers with titanium implants in the upper arm to assess transcutaneous passage.
    • Clinical study of 4 patients with hearing impairment, utilizing titanium implants anchored in the temporal bone for hearing aid attachment.

    Main Results:

    • Implants with designs restricting skin movement showed no inflammatory or adverse soft tissue effects over 7-25 months.
    • Transcutaneous titanium implants for hearing aids remained functional without issues for 38-50 months.
    • Favorable outcomes are attributed to non-alloyed titanium, implant design, and minimally invasive surgery.

    Conclusions:

    • Non-alloyed titanium implants, when properly designed and surgically implemented, can achieve stable and permanent transcutaneous passage.
    • These findings suggest potential for skeletal-anchored external prostheses, such as for facial reconstruction.
    • Further clinical trials are underway to explore these novel treatment possibilities.