Proficiency in testing for hepatitis B surface antigen: experience with Ausria II
Vox Sanguinis
|January 1, 1982
Summary
Lot-to-lot variability in hepatitis B surface antigen (HBsAg) reagents may increase testing errors. Raising the positive to negative (P/N) ratio for validation to 25 is recommended to improve HBsAg test sensitivity and accuracy.
Area of Science:
- * Clinical Chemistry
- * Immunology
- * Public Health
Background:
- * Hepatitis B surface antigen (HBsAg) testing is crucial for blood safety.
- * Radioimmunoassay (RIA) kits were the standard for HBsAg detection in the late 1970s.
- * Ensuring the accuracy and reliability of diagnostic tests is paramount in blood services.
Purpose of the Study:
- * To evaluate the performance of HBsAg proficiency panels using RIA test kits.
- * To identify potential sources of error and variability in HBsAg testing.
- * To recommend improvements for HBsAg test validation and accuracy.
Main Methods:
- * Eight HBsAg proficiency panels were tested.
- * Testing was conducted by routine laboratories in American Red Cross Blood Services Regions.
- * Radioimmunoassay (RIA) test kits were utilized during 1978-1979.
Main Results:
- * Analysis revealed significant lot-to-lot variability in the RIA test kit reagents.
- * This variability was associated with a measurable loss of test sensitivity.
- * An increased error rate in HBsAg detection was observed due to reagent inconsistencies.
Conclusions:
- * Lot-to-lot reagent variability poses a risk to HBsAg testing accuracy.
- * The positive to negative (P/N) ratio is a critical parameter for test validation.
- * Recommending an increased P/N ratio to at least 25 can enhance test sensitivity and reduce errors.