Clinical evaluation of the Scoperimeter, an experimental automatic perimeter
International Ophthalmology
|November 1, 1982
Summary
The Scoperimeter, an automatic campimeter, effectively detects larger visual field defects but misses some early glaucoma. Increasing its resolution could improve early glaucoma detection and assessment.
Area of Science:
- Ophthalmology
- Medical Devices
- Visual Field Testing
Background:
- Automated perimetry is crucial for diagnosing visual field defects.
- The Scoperimeter is an experimental device using a novel suprathreshold strategy.
- Evaluating its efficacy in detecting various visual field abnormalities is necessary.
Purpose of the Study:
- To assess the Scoperimeter's performance in detecting glaucomatous and maculopathy-related visual field defects.
- To compare its diagnostic capabilities against manual perimetry for early glaucoma detection.
Main Methods:
- The Scoperimeter, an automated campimeter, was tested using a gradient-adapted suprathreshold strategy.
- Comparative analysis was conducted on normal visual fields, glaucomatous defects, and maculopathy-related defects.
- Results were compared with high-resolution manual static perimetry.
Main Results:
- The Scoperimeter identified all large visual field defects.
- It detected 60% of very early glaucomatous defects, with manual perimetry identifying the remaining 40%.
- Accurate defect configuration was limited to large, homogeneous defects.
Conclusions:
- The Scoperimeter shows promise for detecting glaucomatous defects, especially with improved resolution and peripheral field examination.
- Its current low resolution and strategy are inadequate for small, central visual field defects.
- Further development is needed for comprehensive visual field assessment.