Related Experiment Videos
Noninvasive thermometry with a clinical x-ray CT scanner
Medical Physics
|September 1, 1982
Summary
Computed Tomography (CT) number shifts can noninvasively measure temperature changes. This study demonstrates CT thermometry
Area of Science:
- Medical Imaging
- Biophysics
- Thermometry
Background:
- CT scanners measure CT numbers proportional to electron density.
- Temperature variations cause density changes, leading to CT number shifts.
- Noninvasive thermometry is crucial for monitoring cancer hyperthermia treatments.
Purpose of the Study:
- To investigate the potential of CT number shifts for noninvasive thermometry.
- To evaluate CT thermometry in vitro for cancer hyperthermia applications.
Main Methods:
- In vitro study using water and muscle tissue samples.
- Utilized an EMI-7070 CT scanner to measure CT numbers.
- Analyzed the relationship between CT numbers, density, and temperature.
Main Results:
- A linear relationship exists between CT number and water density (10-55°C).
- Linear CT number-temperature relationship observed in the hyperthermia range (36-50°C).
- CT number thermal shifts of ~0.4 HU/°C (water) and ~0.45 HU/°C (muscle) were recorded.
Conclusions:
- Thermally generated CT number shifts show promise for noninvasive thermometry.
- Feasible temperature discrimination of a fraction of a degree Celsius is achievable.
- CT thermometry can be applied to monitor temperature during cancer hyperthermia.