Related Experiment Videos
The Ocuplot semi-automatic visual field screener--technical evaluation and clinical results
Documenta Ophthalmologica. Advances in Ophthalmology
|December 1, 1982
Summary
The Ocuplot
Area of Science:
- Ophthalmology
- Medical Device Technology
Background:
- The Ocuplot is a device for visual field testing.
- Previous versions of the Ocuplot have been available.
- The manufacturer claims improvements in newer Ocuplot models.
Purpose of the Study:
- To evaluate the perimetric concept and technical construction of the Ocuplot.
- To assess the efficiency and accuracy of the Ocuplot's test strategy.
Main Methods:
- Evaluation of the mechanical and electrical construction of the Ocuplot.
- Examination of 87 eyes from patients and normal subjects using the Ocuplot.
- Analysis of clinical results from patient examinations.
Main Results:
- The mechanical and electrical construction of the tested Ocuplot was found to be poor.
- The Ocuplot's test strategy is inefficient.
- A high proportion of false positive results were observed during testing.
Conclusions:
- The Ocuplot's current design and test strategy are suboptimal for clinical use.
- Further improvements are needed to enhance the Ocuplot's efficiency and reduce false positives.