Related Experiment Videos
Comparison of electrical field plethysmography with electrical impedance plethysmography
Abstract:
As a means for assessing cardiac function, electrical field plethysmography (EFP) has been shown to have some features quite different from electrical impedance plethysmography (EIP). Here the two techniques are compared by using the two systems simultaneously on a subject and also with independent use in different electrode configurations. The results conform with the view that EIP is related primarily to volumetric changes of the aorta, whereas EFP is affected predominantly by changes in cardiac dimensions and orientation. Because of this difference, the standard time differential formula used for EIP is not applicable for the computation of cardiac output from the EFP waveforms. An alternative method of computation based on the amplitude of the EFP waveform is suggested.