Copper loss from the copper T model TCu22OC

M Thiery, F Schmidt, H J Tatum

    Contraception
    |September 1, 1982
    PubMed
    Summary

    Mineral deposits on Copper T intrauterine devices (IUDs) slow copper release. This study found that increased mineral buildup on IUDs correlates with a reduced copper dissolution rate, impacting device longevity and copper loss.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    c-Rel drives pancreatic cancer metastasis through fibronectin-integrin signaling-induced isolation stress resistance and EMT.

    Molecular cancer·2025
    Same author

    c-Rel drives pancreatic cancer metastasis through Fibronectin-Integrin signaling-induced isolation stress resistance and EMT activation.

    bioRxiv : the preprint server for biology·2025
    Same author

    Bioarchaeology aids the cultural understanding of six characters in search of their agency (Tarquinia, ninth-seventh century BC, central Italy).

    Scientific reports·2024
    Same author

    Treatment of Olfactory Disorders After SARS - CoViD 2 Virus Infection.

    Ear, nose, & throat journal·2023
    Same author

    Postoperative serum C-reactive protein dynamics after pharyngolaryngectomy with jejunal free-flap reconstruction.

    Annals of the Royal College of Surgeons of England·2022
    Same author

    Martian CO<sub>2</sub> Ice Observation at High Spectral Resolution With ExoMars/TGO NOMAD.

    Journal of geophysical research. Planets·2022

    Area of Science:

    • Reproductive Health
    • Biomaterials Science
    • Medical Device Analysis

    Background:

    • Copper T intrauterine devices (IUDs) are widely used for long-term contraception.
    • The presence of mineral deposits on IUDs is a known phenomenon, but its effect on copper release is not fully understood.

    Purpose of the Study:

    • To investigate the impact of mineral deposits on the copper dissolution rate of the Copper T Model TCu22OC.
    • To correlate mineral deposit characteristics with copper loss over time.

    Main Methods:

    • Examination of 118 explanted Copper T22OC IUDs (12-72 months in utero).
    • Grading of mineral deposit coverage on the copper surface.
    • Copper analysis of 30 devices to determine daily copper loss.
    Keywords:
    BelgiumBleedingContraceptionContraception TerminationContraceptive MethodsCopper--analysisDeveloped CountriesEuropeExaminations And DiagnosesFamily PlanningIngredients And ChemicalsInorganic ChemicalsIudIud MaterialIud, Copper ReleasingLaboratory Examinations And DiagnosesLaboratory ProceduresMetalsPainPregnancyPregnancy, PlannedPregnancy, UnplannedSizeWestern Europe

    Related Experiment Videos

    Main Results:

    • Mineral deposits increased in size during the first three years of IUD use.
    • A mean daily copper loss of 29.4 micrograms was observed (range: 7.3-62.1 micrograms).
    • An inverse correlation was found between mineral deposit amount and copper loss rate.

    Conclusions:

    • Mineral deposits significantly influence the copper dissolution rate of Copper T IUDs.
    • Increased mineral deposition on the copper surface reduces the rate of copper loss.