Improved ear oximetry with a rubefacient ointment

A J Schneider, R Miller, W R Baetz

    Critical Care Medicine
    |November 1, 1978
    PubMed
    Summary

    Topical ointment improves noninvasive ear oximetry accuracy in cardiac surgery patients. This method enhances arterial oxygen saturation measurements, especially in patients with poor peripheral circulation.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    [Tests for alcohol consumption during pregnancy: what biomarkers are suitable?].

    Nederlands tijdschrift voor geneeskunde·2016
    Same author

    The feasibility of incorporating structured therapeutic consultations with real patients into the clinical clerkship internal medicine.

    Naunyn-Schmiedeberg's archives of pharmacology·2012
    Same author

    A context-learning pharmacotherapy program for preclinical medical students leads to more rational drug prescribing during their clinical clerkship in internal medicine.

    Clinical pharmacology and therapeutics·2009
    Same author

    The anatomical limits of the posterior vaginal vault toward its use as route for intra-abdominal procedures.

    Surgical endoscopy·2008
    Same author

    The relationship between arterial PO2 and mixed venous PO2 in response to changes in positive end-expiratory pressure in ventilated patients.

    Anaesthesia·2008
    Same author

    [First abdominal access in gynecological laparoscopy: which method to utilize?].

    Minerva ginecologica·2006

    Area of Science:

    • Medical Devices
    • Cardiovascular Surgery
    • Anesthesiology

    Background:

    • Accurate arterial oxygen saturation (SpO2) monitoring is critical in postoperative cardiac surgery patients.
    • Peripheral vasoconstriction can impair noninvasive oximetry readings in these patients.
    • Ear oximetry offers a potential alternative but requires adequate ear blood flow.

    Purpose of the Study:

    • To evaluate the efficacy of a rubefacient ointment in improving noninvasive ear oximetry accuracy.
    • To compare ear oximetry readings with arterial blood gas (ABG) measurements.
    • To assess the utility of ear oximetry in patients with potentially compromised peripheral circulation.

    Main Methods:

    • 51 adult cardiac surgery patients were studied in the immediate postoperative period.
    • Ear arterialization was achieved via rubbing (control) or topical rubefacient ointment (experimental).
    • Oximeter readings from both ears were compared to ABG-derived saturation values.

    Main Results:

    • Untreated ear oximetry readings were significantly lower than ABG values (p < 0.001).
    • Ointment-treated ear oximetry readings showed no significant difference from ABG values (0.05 < p < 0.10).
    • The rubefacient improved ear blood flow, enhancing oximetry accuracy.

    Conclusions:

    • Topical rubefacient ointment significantly improves the accuracy of noninvasive ear oximetry.
    • This technique is particularly beneficial for patients with cold extremities or peripheral vasoconstriction.
    • Ear oximetry, enhanced by rubefacient, provides a reliable method for SpO2 monitoring in challenging postoperative conditions.