Tonometry in glaucomatous globes

K N Gelatt, G G Gum, K P Barrie

    Investigative Ophthalmology & Visual Science
    |May 1, 1981
    PubMed
    Summary

    The Mackay-Marg and pneumatonograph tonometers are reliable for measuring intraocular pressure in glaucomatous beagles, even at high pressures. The EMT-20 tonometer showed poor reliability in this study.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Autosomal recessive primary open angle glaucoma (POAG) in beagles is not associated with mutations in the myocilin (MYOC) gene.

    Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie·2009
    Same author

    Deep lamellar endothelial keratoplasty in 10 horses.

    Veterinary ophthalmology·2008
    Same author

    Intranictitans tacking for replacement of prolapsed gland of the third eyelid in dogs.

    Veterinary ophthalmology·2008
    Same author

    Myocilin protein levels in the aqueous humor of the glaucomas in selected canine breeds.

    Veterinary ophthalmology·2008
    Same author

    Aqueous humor myocilin protein levels in normal, genetic carriers, and glaucoma Beagles.

    Veterinary ophthalmology·2008
    Same author

    Corneal transplantation for inflammatory keratopathies in the horse: visual outcome in 206 cases (1993-2007).

    Veterinary ophthalmology·2008

    Area of Science:

    • Ophthalmology
    • Veterinary Medicine
    • Biomedical Engineering

    Background:

    • Glaucoma is a leading cause of irreversible blindness.
    • Accurate intraocular pressure (IOP) measurement is crucial for glaucoma diagnosis and management.
    • Evaluating tonometer performance in animal models is essential for understanding their clinical utility.

    Purpose of the Study:

    • To assess the reliability of the Mackay-Marg, pneumatonograph, and EMT-20 tonometers for measuring IOP in glaucomatous beagle eyes.
    • To compare tonometer readings with direct manometric measurements of anterior chamber pressure.

    Main Methods:

    • Three tonometers (Mackay-Marg, pneumatonograph, EMT-20) were tested on glaucomatous beagle eyes.
    • Tonometry was performed with the manometric system both closed and open to a reservoir column.
    • Tonometer readings were correlated with manometric recordings to determine reliability (r2 values).

    Main Results:

    • The Mackay-Marg tonometer demonstrated high reliability (r2 = 0.852 closed, 0.816 open) up to 50 mm Hg, with slightly lower reliability (r2 = 0.818 closed, 0.740 open) up to 100 mm Hg.
    • The pneumatonograph also showed high reliability (r2 = 0.892 closed, 0.806 open) up to 50 mm Hg and good reliability (r2 = 0.796 closed, 0.75 open) up to 100 mm Hg.
    • The EMT-20 tonometer exhibited poor reliability, with r2 values ranging from 0.235 to 0.654.

    Conclusions:

    • The Mackay-Marg and pneumatonograph tonometers are reliable tools for IOP measurement in glaucomatous beagle eyes, even at elevated pressures.
    • The EMT-20 tonometer is not recommended for use in glaucomatous eyes due to its unsatisfactory performance.
    • This study highlights the importance of device selection for accurate IOP assessment in veterinary ophthalmology.