Related Experiment Videos
Transcutaneous oxygen measurement during thoracic anaesthesia
Acta Anaesthesiologica Scandinavica
|December 1, 1980
Summary
Transcutaneous oxygen tension (tcPO2) can serve as a useful oxygen parameter alongside arterial oxygen tension (PaO2) during thoracic anesthesia. However, tcPO2 interpretation requires a preoperative baseline measurement for accurate comparison.
Area of Science:
- Anesthesiology
- Respiratory Physiology
- Medical Monitoring
Background:
- Monitoring oxygenation is crucial during anesthesia.
- Transcutaneous oxygen tension (tcPO2) offers a non-invasive method for assessing tissue oxygenation.
- Its utility during thoracic anesthesia requires further investigation.
Purpose of the Study:
- To evaluate the value of tcPO2 as an oxygen parameter during uncomplicated thoracic anesthesia.
- To compare tcPO2 measurements with arterial oxygen tension (PaO2) under various inspiratory oxygen fractions (FIO2).
- To assess potential interference of anesthetic agents with tcPO2 readings.
Main Methods:
- Ten patients undergoing thoracic anesthesia with oxygen-nitrous oxide and enflurane or flunitrazepam/fentanyl were studied.
- tcPO2 was measured using a Radiometer TCM-I monitor at 45°C.
- Measurements were compared with simultaneously obtained PaO2 values preoperatively and peroperatively.
Main Results:
- No significant interference from anesthetic agents was observed.
- tcPO2 was consistently lower than PaO2 across all tested FIO2 levels.
- No statistically significant differences were found between PaO2 and tcPO2 at FIO2 of 0.5, 0.4, and 0.3 when comparing peroperative to preoperative values.
- Linear regression revealed a disparity between pre- and peroperative PaO2/tcPO2 ratios, indicating a decreasing ratio with increasing PaO2.
Conclusions:
- tcPO2 cannot fully substitute for PaO2 during thoracic anesthesia.
- Both tcPO2 and PaO2 demonstrated comparable utility as oxygen parameters in this patient group.
- Accurate interpretation of peroperative tcPO2 necessitates a preoperative measurement as a reference point.