Abstract:
A new carbon transcutaneous access device (CTAD) for hemodialysis id described the precludes the need for needle puncture. The device consists of a vitreous carbon access port attached to a PTFE graft. A disposable connector provides for movement of blood from the device into and out of the dialyzer. Twenty-one of the devices have been implanted in 18 patients. Overall 9 month patency rate is 64.3%, comparing favorably with conventional PTFE grafts. The incidence of infection (1 per .05 patient months) and thrombosis (28.6% at 9 months) similarly compare favorably with other forms of vascular access. The CTAD provides a unique opportunity by permitting hemodialysis without the pain or risk of needle punctures.