Summary
Many fluoroscopic systems exceed recommended radiation exposure limits, with 42% measuring above 5 R/min. This study highlights the need for improved radiation protection measures in medical imaging.
Area of Science:
- Medical Physics
- Radiological Health
Context:
- Assessment of fluoroscopic system radiation exposure rates.
- Evaluation of 155 fluoroscopic systems over two years.
Purpose:
- To measure and analyze exposure rates of fluoroscopic systems.
- To identify systems exceeding recommended radiation protection limits.
Summary:
- 42% of surveyed fluoroscopic systems exceeded the 5 R/min radiation exposure limit.
- Systems with television links generally showed lower exposure rates.
- Significant variations in exposure rates were noted between different manufacturers' equipment.
Impact:
- Identified non-compliance with radiation protection codes.
- Recommendations for corrective measures to reduce population radiation dose.
- Potential for significant radiation dose reduction in Quebec through improved practices.