Effect of bentonites on fluorometric selenium determination
Summary
Bentonites interfere with selenium analysis in animal feed. A simple dilution and centrifugation method prevents this error, ensuring accurate selenium (Se) measurements in feed samples.
Area of Science:
- Analytical Chemistry
- Animal Nutrition
- Trace Element Analysis
Background:
- Accurate quantification of selenium (Se) is crucial for animal feed analysis.
- The AOAC fluorometric method (3.097-3.101) is a standard technique for Se determination.
- Bentonites, commonly used as feed additives, have been observed to interfere with this analysis.
Purpose of the Study:
- To identify the cause of interference by bentonites in the AOAC fluorometric method for Se analysis.
- To develop a modified procedure to overcome this interference and ensure accurate Se measurements.
Main Methods:
- Investigated the interaction between bentonite's insoluble residue and the piazselenol intermediate.
- Implemented a modified analytical procedure involving dilution, reduction with HCl, and centrifugation.
- Analyzed the clear supernatant liquid for accurate Se quantification.
Main Results:
- Bentonite's insoluble residue binds the piazselenol, leading to underestimation of Se.
- The modified method, utilizing dilution and centrifugation, effectively removes the interfering residue.
- Accurate Se analysis is achievable even in the presence of bentonite using the proposed technique.
Conclusions:
- Bentonite-induced errors in Se analysis can be reliably avoided by employing a simple dilution and centrifugation step.
- This modification ensures the integrity of the AOAC fluorometric method for Se determination in feeds containing bentonites.
- The findings provide a practical solution for accurate selenium monitoring in animal nutrition.