The environmental density artifact: a beam-hardening effect in computed tomography
Radiology
|October 1, 1981
Summary
Beam hardening in computed tomography (CT) causes an environmental density artifact, altering CT numbers in central regions. This effect, observed in phantom studies, can lead to misdiagnosis during contrast-enhanced scans.
Area of Science:
- Medical Imaging Physics
- Radiological Sciences
Background:
- Computed tomography (CT) imaging relies on accurate attenuation measurements.
- Beam hardening is a known phenomenon in CT, causing energy-dependent X-ray attenuation.
- Environmental density artifacts can affect CT number accuracy, particularly in heterogeneous regions.
Purpose of the Study:
- To quantify the environmental density artifact in computed tomography.
- To investigate the impact of varying attenuation coefficients on central CT numbers.
- To assess the potential diagnostic implications of this artifact during contrast-enhanced CT.
Main Methods:
- Concentric cylindrical phantoms were utilized for measurements.
- The outer cylinder contained solutions with different attenuation coefficients (iodinated contrast medium, isopropyl alcohol).
- The central cylinder was filled with water to measure CT number changes.
Main Results:
- A measurable change in the CT number of the central water cylinder was observed.
- When the outer cylinder's CT number increased by approximately 100 HU, the central water cylinder's CT number increased by about 5 HU.
- This indicates a direct relationship between the surrounding material's attenuation and the central region's measured density.
Conclusions:
- The environmental density artifact, stemming from beam hardening, demonstrably alters CT numbers.
- This artifact can introduce inaccuracies in CT imaging, especially when contrast enhancement is used.
- Clinicians should be aware of this artifact to avoid potential misdiagnosis in contrast-enhanced CT examinations.