Automated platelet counting--a re-evaluation of the sedimentation method
Clinical and Laboratory Haematology
|January 1, 1981
Summary
Investigating the Coulter Thrombocounter, this study found that sample removal from the sedimented plasma layer causes platelet count discrepancies. Strict adherence to manufacturer instructions generally yields satisfactory results, though higher counts may occur above 200 X 10(9)/l.
Area of Science:
- Hematology
- Laboratory Medicine
- Clinical Pathology
Background:
- Interlaboratory quality assessment surveys highlight discrepancies in platelet counting.
- The Coulter Thrombocounter with Thrombofuge method is widely used for platelet enumeration.
Purpose of the Study:
- To identify the causes of discrepancies in platelet counts obtained using the Coulter Thrombocounter with Thrombofuge method.
- To evaluate the accuracy, precision, drift, and carry-over of the Thrombocounter method.
Main Methods:
- Evaluation of the Coulter Thrombocounter using the Thrombofuge method for specimen preparation.
- Assessment of accuracy, precision, drift, and carry-over.
- Comparison with a whole blood reference method.
Main Results:
- Discrepancies in platelet counts were observed during routine use.
- The level of the sedimented plasma layer significantly contributed to errors.
- Strict adherence to manufacturer's instructions improved results, making them comparable to the reference method.
- The Thrombofuge method tended to yield higher platelet counts than the reference method for counts exceeding 200 X 10(9)/l, independent of other factors.
Conclusions:
- The sedimented plasma layer is a critical factor influencing platelet count accuracy with the Thrombofuge method.
- Standardized use according to manufacturer guidelines generally ensures reliable platelet counts.
- A systematic overestimation of platelet counts above 200 X 10(9)/l may occur with this method.