Non-specific count subtraction in radioimmunoassay (a criticism)
Summary
The radioimmunoassay (RIA) blank value does not accurately reflect nonspecific radioactivity. Subtracting this blank value from RIA results leads to analytical errors, as demonstrated by Gastrin-RIA studies.
Area of Science:
- Immunology
- Analytical Chemistry
- Biochemistry
Background:
- Radioimmunoassay (RIA) blank values are classically determined but may not accurately represent nonspecific radioactivity.
- The composition of blank tubes differs significantly from standard and unknown tubes in RIA assays.
Purpose of the Study:
- To provide evidence that subtracting the blank value (Blk) from RIA results is analytically incorrect.
- To evaluate the impact of the blank subtraction procedure on RIA data accuracy using Gastrin-RIA as a model.
Main Methods:
- Gastrin-RIA with charcoal adsorption for bound-free separation was used.
- Antigen-recovery and parallelism (serum dilution) tests were performed.
- Results from blank subtraction (BS) values were compared with non-blank subtraction (BNS) values.
Main Results:
- Blank subtraction (BS) values in recovery tests were higher than expected, unlike quantitative non-blank subtraction (BNS) values.
- Serum dilution tests showed BNS values yielded curves more parallel to the BNS standard curve than BS values did to the BS standard curve.
Conclusions:
- The classical blank subtraction procedure in RIA is analytically incorrect and should be avoided.
- Accurate RIA data processing requires omitting the empirical blank value subtraction.