Self-monitoring of blood glucose without a meter
We studied the reliability of the Chemstrip system in the self-monitoring of blood glucose by diabetic patients. A total of 453 blood glucose values obtained by 38 diabetic patients using Chemstrip were compared with the glucose values of the same blood samples measured with a conventional laboratory method. We found a definite tendency at concentrations less than 320 mg/dl for patient-read Chemstrip values to be too low. Also, we found a wide range in laboratory values corresponding to a given patient-read Chemstrip value. However, a better agreement between Chemstrip and laboratory values was obtained when the Chemstrips were read by trained staff. We recommend, therefore, that the use of Chemstrip in the self-monitoring of blood glucose can be restricted to patients who have shown competence in the use of Chemstrip during prior testing.
We studied the reliability of the Chemstrip system in the self-monitoring of blood glucose by diabetic patients. A total of 453 blood glucose values obtained by 38 diabetic patients using Chemstrip were compared with the glucose values of the same blood samples measured with a conventional laboratory method. We found a definite tendency at concentrations less than 320 mg/dl for patient-read Chemstrip values to be too low. Also, we found a wide range in laboratory values corresponding to a given patient-read Chemstrip value. However, a better agreement between Chemstrip and laboratory values was obtained when the Chemstrips were read by trained staff. We recommend, therefore, that the use of Chemstrip in the self-monitoring of blood glucose can be restricted to patients who have shown competence in the use of Chemstrip during prior testing.