Serum gentamicin assay by a radiometric procedure
American Journal of Clinical Pathology
|February 1, 1980
Summary
A new BACTEC radiometric assay for gentamicin serum levels is fast and reproducible. However, it shows significant differences at concentrations above 4 microgram/ml compared to other methods.
Area of Science:
- Clinical Chemistry
- Microbiology
- Pharmacokinetics
Background:
- Accurate determination of gentamicin serum levels is crucial for therapeutic drug monitoring.
- Existing microbiologic assays and radioimmunoassay (RIA) methods have limitations in speed or cost.
- Radiometric methods offer potential for rapid and sensitive drug level determination.
Purpose of the Study:
- To evaluate a new radiometric (BACTEC) assay for gentamicin serum level determination.
- To compare the BACTEC assay's performance against established microbiologic assays and RIA.
- To assess the accuracy, speed, and reproducibility of the BACTEC method.
Main Methods:
- A 14C-urea substrate and Proteus species culture were used in the BACTEC assay.
- Gentamicin levels in serum were determined using the BACTEC assay, three non-radiometric microbiologic assays, and RIA.
- Assay performance was compared across different gentamicin concentration ranges.
Main Results:
- The BACTEC assay was found to be simple, fast, and reproducible.
- At gentamicin concentrations < 4 microgram/ml, BACTEC accuracy was comparable to other assays.
- At gentamicin concentrations > 4 microgram/ml, BACTEC showed significant differences, reading lower than RIA and other assays.
Conclusions:
- The BACTEC assay is a viable alternative to RIA for laboratories with radiometric equipment, especially for batch processing.
- Due to cost, BACTEC is less practical than rapid, inexpensive microbiologic assays for routine gentamicin monitoring.
- The BACTEC assay's accuracy is concentration-dependent, requiring careful consideration for therapeutic drug monitoring.