Abstract:
We tested the response of ion-selective sodium and potassium electrodes that are a part of the new multi-channel chemical analyzers, Astra 8 and Astra 4 (Beckman). For this we used plain and albumin-containing aqueous solutions of known Na+ and K+ concentrations, varied inversely so we could assess how the concentration of each electrolyte would affect the response of the other electrode. The stability of the electrodes with time was good and the electrodes are indeed selective for Na+ and K+.
Related Concept Videos
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...