CT quality assurance: computer assisted slice thickness determination

N J Schneiders, S C Bushong

    Medical Physics
    |January 1, 1980
    PubMed
    Summary
    This summary is machine-generated.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    In x-ray computed tomography, technique factors should be selected appropriate to patient size.

    Medical physics·2001
    Same author

    History of standards, certification, and licensure in medical health physics.

    Health physics·1995
    Same author

    Hazards evaluation of neuroangiographic procedures.

    AJNR. American journal of neuroradiology·1994
    Same author

    Exposure rates during special procedures with C-arm type fluoroscopic systems.

    Radiology·1994
    Same author

    The pregnant technologist and magnetic resonance imaging.

    Radiology management·1993
    Same author

    What to look for when buying a mammo system.

    Diagnostic imaging·1992
    Same journal

    Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

    Medical physics·2026
    Same journal

    Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

    Medical physics·2026
    Same journal

    Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

    Medical physics·2026
    Same journal

    Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

    Medical physics·2026
    Same journal

    Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

    Medical physics·2026
    Same journal

    Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

    Medical physics·2026
    See all related articles

    Accurate computed tomography (CT) slice thickness is crucial for image quality. Automated analysis of the AAPM phantom revealed significant discrepancies between CT scanner settings and actual slice thicknesses, highlighting the need for routine quality assurance.

    Area of Science:

    • Medical Imaging Physics
    • Radiological Technology
    • Quality Assurance in Medical Imaging

    Background:

    • Precise slice geometry is a critical but challenging parameter to determine for CT scanners.
    • Slice thickness directly impacts image quality and diagnostic accuracy in CT scans.

    Purpose of the Study:

    • To develop and evaluate a series of computer programs for automated analysis of CT slice thickness.
    • To assess the accuracy of slice thickness measurements using the AAPM phantom and specific CT scanners.

    Main Methods:

    • Utilized a dedicated slice thickness insert within the AAPM phantom.
    • Developed computer programs to automatically generate beam profiles and measure slice thicknesses at multiple points.
    • Performed analysis on an AS&E scanner with various slice thickness settings (2-10 mm).

    Related Experiment Videos

    Main Results:

    • Automated analysis successfully generated beam profiles and slice thickness data without operator intervention.
    • Measured slice thicknesses varied significantly from the scanner's nominal settings, ranging from 3 mm to 8 mm for settings of 2 mm to 10 mm.
    • Identified discrepancies between programmed and actual slice thicknesses across different settings.

    Conclusions:

    • The developed automated analysis provides an efficient method for assessing CT slice geometry.
    • Significant variations in measured slice thickness underscore the necessity of regular quality assurance protocols.
    • Implementing routine slice thickness measurements is essential for maintaining diagnostic image quality and patient safety.