Stability of international reference thromboplastins
British Journal of Haematology
|May 1, 1980
Summary
The long-term stability of thromboplastin preparations was assessed over a decade. Results show no significant changes in clotting times, ensuring reliable diagnostic use of these WHO International Reference Preparations.
Area of Science:
- Clinical Chemistry
- Biotechnology
- Medical Diagnostics
Background:
- The WHO International Reference Preparation (IRP) of thromboplastin is crucial for standardizing coagulation tests.
- Assessing the long-term stability of reference materials is essential for maintaining the accuracy and reliability of diagnostic assays.
Purpose of the Study:
- To evaluate the long-term stability of the WHO IRP 67/40 and two secondary preparations (WHO IRP 68/434 and Dutch RP 73/528) over a 10-year period.
- To determine if any changes in the preparations' performance would impact clinical laboratory results.
Main Methods:
- Clotting times of normal plasmas were measured using the reference preparations between 1969 and 1979.
- Linear regression analysis was employed to assess the slopes of lines relating clotting-time ratios.
- Data were corrected for modifications in assay techniques over the study period.
Main Results:
- Clotting times determined with the WHO IRP 67/40, WHO IRP 68/434, and Dutch RP 73/528 showed no practically significant changes over the 10-year period.
- Linear regression analysis did not reveal any systematic changes in the slopes of the lines relating clotting-time ratios over time.
- The stability of these thromboplastin preparations was confirmed, indicating consistent performance.
Conclusions:
- The WHO International Reference Preparation (IRP) of thromboplastin 67/40 and the secondary preparations (WHO IRP 68/434 and Dutch RP 73/528) demonstrated excellent long-term stability.
- These findings support the continued reliable use of these preparations in standardizing prothrombin time assays.
- The stability ensures consistent diagnostic accuracy in clinical laboratories utilizing these reference materials.