Related Experiment Videos
Determination of avermectins in plasma at nanogram levels using high-performance liquid chromatography with
Journal of Chromatography
|April 4, 1980
Abstract:
An analytical method is described for the determination of the avermectins in plasma based upon high-performance liquid chromatography of fluorescent derivatives of these compounds. The analyte is isolated by adsorption chromatography on Florisil, dehydrated in an acetic anhydride-pyridine mixture, and the fluorophore is further separated by chromatography on silica gel in advance of introduction into a reversed-phase system. This method, which can be applied to samples containing as little as 0.2 ng drug per ml, has an accuracy of 5% mean relative error and a precision of 8% relative standard deviation. A study and discussion of several factors which affect the analytical reaction are included.