Summary
Microwave leakage during patient treatment is higher than predicted by phantom tests. This finding is crucial for ensuring safety during microwave hyperthermia therapy.
Area of Science:
- Biomedical Engineering
- Medical Physics
- Radiation Oncology
Background:
- The Bureau of Radiological Health (BRH) proposed a draft standard for microwave leakage at 10 mW/cm2.
- Accurate measurement of microwave leakage is essential for patient safety during therapeutic applications.
Purpose of the Study:
- To compare microwave leakage measurements under static phantom conditions versus during actual patient treatments.
- To assess if current phantom-based measurements accurately predict real-world microwave exposure.
Main Methods:
- Microwave leakage was measured using a tissue-equivalent phantom, varying net forward power and aperture-to-skin distance (SSD).
- Leakage was also measured on patients undergoing microwave hyperthermia treatments at various anatomical locations.
Main Results:
- Microwave leakage during patient treatments was significantly higher than predicted by static phantom measurements.
- Anatomical site and treatment conditions appear to influence the extent of microwave leakage.
Conclusions:
- Static phantom models may underestimate microwave leakage during clinical hyperthermia.
- Further research is needed to refine safety standards and measurement protocols for microwave hyperthermia to account for in-vivo conditions.