Air-radiated sound from bone vibration transducers and its implications for bone conduction audiometry
British Journal of Audiology
|August 1, 1980
Summary
Air-borne sound from bone conduction audiometry vibrators can affect hearing tests. Using earplugs at 3 and 4 kHz minimizes errors for vibrator types B70A, B71, and B72.
Area of Science:
- Audiology
- Acoustics
- Hearing Science
Background:
- Air-borne sound leakage from bone conduction audiometry vibrators is a known issue.
- Quantifying this effect is crucial for accurate hearing threshold measurements.
Purpose of the Study:
- To quantify the air-borne sound radiation from B70A, B71, and B72 bone conduction vibrator types.
- To assess the impact of this sound on hearing threshold measurements.
Main Methods:
- Sound pressure levels were measured at the ear canal entrance for three examples of each vibrator type on four human heads.
- Measured values were compared to hearing threshold levels using two methods: international standards and direct subject measurements.
Main Results:
- None of the tested vibrators fully met international standards for air-borne sound radiation across all frequencies.
- Significant errors in threshold measurements are unlikely at 3 and 4 kHz when a high-quality earplug is used.
Conclusions:
- Bone conduction vibrator air-borne sound can influence audiometric results.
- Proper use of earplugs, particularly at 3 and 4 kHz, can mitigate potential measurement errors.