Evaluation of automatic and semiautomatic coagulation assay instruments
Clinical and Laboratory Haematology
|January 1, 1980
Summary
Instrument precision for activated partial thromboplastin time (PTT) and prothrombin time (PT) assays varied significantly across devices and reagents. The Coagulyzer, Coag-A-Mate, and Auto-Fi showed the best precision for PTT assays.
Area of Science:
- Clinical Chemistry
- Hematology
- Medical Laboratory Science
Background:
- Activated partial thromboplastin time (PTT) and prothrombin time (PT) are crucial diagnostic assays.
- Instrument and reagent variability can impact the accuracy and reliability of coagulation test results.
- Standardization of coagulation testing is essential for consistent patient care.
Purpose of the Study:
- To compare the precision of five different hemostasis instruments in performing PTT assays.
- To evaluate the impact of three distinct coagulation reagent systems on PTT assay performance.
- To assess the correlation and agreement between instruments for both PT and PTT assays.
Main Methods:
- Five instruments (Coag-A-Mate, Auto-Fi, Coagulyzer, Electra 600, Fibrometer) were evaluated.
- Three reagent systems (Ortho, Dade, General Diagnostics) were used with normal and abnormal controls.
- PT and PTT assays were performed on healthy donors and patient samples.
Main Results:
- Significant variations in mean PTT values and coefficients of variation (CV) were observed across instruments and reagents.
- The Coagulyzer (Dade), Coag-A-Mate (Dade, General Diagnostics), and Auto-Fi (General Diagnostics) demonstrated the lowest CVs for PTT.
- Poor correlation was found for PT results, while PTT results between Coag-A-Mate and Auto-Fi showed a correlation of 0.84.
- Coincidence rates for classifying patient samples as normal/abnormal were 91.2% for PTT and 85.1% for PT.
Conclusions:
- Instrument and reagent choice significantly affects PTT assay precision.
- The Coagulyzer, Coag-A-Mate, and Auto-Fi offer superior precision for PTT testing under specific reagent conditions.
- Discrepancies in PT and PTT classifications highlight the need for careful instrument selection and method validation.