The focus-to-detector distance as a source of systematical errors in the measurement of Chaoul therapy units
Abstract:
The skin exposure rates measured on 22 Chaoul units in two consecutive years were compared and their variance was analysed. The statistical fluctuation of the ionization method was 3.1% by a factor of about 2 to 2.5 smaller than the variations due to lack of reproducibility of the Chaoul units. The authors observed systematical errors among exposure rate measurement performed at different focus-to-detector distances. The effective source-to-detector distance is different for various ionization chambers. It is the sum of nominal focus-to-detector distance plus a geometrical constant. The geometrical constant is for a particular chamber only to a small extent dependent on the front wall thickness and on the focus-to detector distance. Sufficient standardization of both calibration procedure and construction of ionization chambers may help in avoiding this effect.
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Distance Corrections
Common Leveling Mistakes and Errors