An improved signal processor for the ultrasonic dimension gauge
Cardiovascular Research
|August 1, 1980
Summary
A new sonomicrometer design prevents reset errors caused by transmission artifacts, enabling accurate ultrasonic measurement of aortic diameter in dogs.
Area of Science:
- Biomedical Engineering
- Cardiovascular Physiology
Background:
- Sonomicrometers use ultrasonic transit time for diameter measurement.
- Transmission artifacts frequently caused errors in previous sonomicrometer designs, leading to inaccurate readings.
Purpose of the Study:
- To improve sonomicrometer design by eliminating reset errors caused by transmission artifacts.
- To develop a reliable method for monitoring ascending aortic diameter.
Main Methods:
- Modified the sonomicrometer circuit by incorporating two monostable vibrators and one NAND gate.
- Tested the improved design by monitoring ascending aortic diameter in unanaesthetised dogs over 6-8 weeks.
Main Results:
- The enhanced design successfully corrected resetting errors caused by transmission artifacts.
- The new flip-flop pulse accurately reflected the ultrasonic burst transit time through the aorta.
Conclusions:
- The improved sonomicrometer design provides a reliable and artifact-free method for measuring aortic diameter.
- This technology facilitates long-term, non-invasive monitoring of cardiovascular parameters in animal models.