Influence of expanded balloon diameter on Palmaz-Schatz stent recoil
E Rechavia1, F Litvack, G Macko
1Department of Medicine, Cedars-Sinai Medical Center, Los Angeles, California 90048, USA.
Abstract:
After successful stent implantation, the residual luminal diameter of the stented vessel is usually smaller than the maximal stent-expanded balloon diameter. The goal of this study was to determine whether immediate vessel diameter recoil after Palmaz-Schatz stenting is affected by the final expanding balloon diameter used during stent deployment. Single Palmaz-Schatz balloon expandable stents were successfully placed in 108 stenotic lesions. There were 68 patients with 75 saphenous vein graft (SVG) and 30 patients with 33 native coronary artery lesions, including 26 restenotic and 82 de novo occlusive (> 50% diameter stenosis) lesions. Quantitative coronary angiography was used for the assessment of stent recoil, defined as the difference between the minimal diameter of the fully expanded balloon and the postprocedure minimal lumen diameter divided by minimal diameter of the fully expanded balloon. A strong correlation (r = 0.94) was found between the minimal diameter of the fully expanded balloon and poststenting minimal lumen diameter. Immediate recoil was 11.3 +/- 7.5%, responsible on an average for 0.4 +/- 0.2-mm acute lumen loss. Recoil was less in SVG than in coronary arteries (9.7 +/- 6.6% vs. 14.0 +/- 7.8%; P = 0.004, and 0.3 +/- 0.2 vs. 0.4 +/- 0.2 mm; p = 0.01). Lesions were divided into four subgroups, based on the final stent expanding balloon diameter: (1) < or = 3.0 mm (n = 33); (2) > 3 < or = 3.5 mm (n = 43); (3) > 3.5 < or = 4 mm (n = 23); and (4) > 4 mm (n = 9).(ABSTRACT TRUNCATED AT 250 WORDS)