Estimation of bladder volume using portable ultrasound in clam enterocystoplasty patients
J W Barrington1, A Jones, J Robinson
1Department of Urology, Cardiff Royal Infirmary, United Kingdom.
Purpose:
We evaluated portable ultrasound for the measurement of residual volume following clam enterocystoplasty.
Materials And Methods:
Bladder volumes were measured in 25 clam enterocystoplasty patients using a portable ultrasound machine and compared with urethral catheter volumes. A similar study was performed on a control group of 50 patients with normal bladder architecture.
Results:
There was a close correlation (r = 0.978) between ultrasound and catheter volumes in the control group with a regression line slope of 0.88. There was an equally close correlation (r = 0.960) in the clam enterocystoplasty group but the regression line slope was lower at 0.65, which suggests an underestimation of the true bladder volume. There was a significant difference (p < 0.0001) between the ultrasound-to-catheter volume ratio in the 2 groups.
Conclusions:
Portable ultrasound is inaccurate for the measurement of residual bladder volume in clam enterocystoplasty patients.