Influence of internal current and pacing current on pacemaker longevity

A Schuchert1, K H Kuck

  • 1Department of Cardiology, University-Hospital Eppendorf, Hamburg, Germany.

Pacing and Clinical Electrophysiology : PACE
|January 1, 1994
PubMed
Summary

Programming VVI pacemakers to lower pulse amplitudes, like 2.5 V, significantly extends battery longevity. The newer Minix pacemaker showed greater longevity improvements compared to the older Pasys model, especially at reduced pulse amplitudes.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Optimising access: safety and efficacy of venous closure devices in cryoballoon pulmonary vein isolation.

Frontiers in cardiovascular medicine·2025
Same author

Predictors of Atrial Fibrillation Recurrence after Catheter Ablation: Data from the German Ablation Registry.

Scientific reports·2017
Same author

[Cardiac arrest in a fitness trainer with apical hypertrophic cardiomyopathy associated with cor triatriatum sinister].

Medizinische Klinik, Intensivmedizin und Notfallmedizin·2017
Same author

[Brain infarction as initial manifestation of eosinophilic myocarditis].

Der Nervenarzt·2017
Same author

Safety and efficacy of the percutaneous transaxillary access for transcatheter aortic valve implantation using various transcatheter heart valves in 100 consecutive patients.

International journal of cardiology·2017
Same author

The effects of local steroid-elution on high-performance pacing leads.

Herzschrittmachertherapie & Elektrophysiologie·2016

Area of Science:

  • Biomedical Engineering
  • Medical Device Technology

Background:

  • Pacemaker longevity is crucial for patient care and device management.
  • Comparing new and older pacemaker models helps understand technological advancements.
  • Battery current consumption directly impacts pacemaker lifespan.

Purpose of the Study:

  • To evaluate the effect of reduced pulse amplitude on battery current and longevity.
  • To compare a new VVI pacemaker (Minix 8341) with a former model (Pasys 8329).

Main Methods:

  • Telemetric measurement of battery current at various pulse amplitudes and durations.
  • Calculation of pacing current by subtracting internal current from battery current.
  • Projecting pacemaker longevity based on measured currents and programming settings.

Related Experiment Videos

Main Results:

  • The Minix pacemaker exhibited significantly lower battery current due to lower internal current.
  • Programming from 5.0 V to 2.5 V pulse amplitude reduced battery current more effectively in the Minix.
  • Programming to 2.5 V extended projected longevity by 2.3 years for Pasys and 7.1 years for Minix.

Conclusions:

  • Pacemaker longevity can be extended by programming to 2.5 V or lower.
  • This strategy is particularly effective for pacemakers requiring low internal current.