Molecular profiling and weight determination of heparins and depolymerized heparins
A Ahsan1, W Jeske, D Hoppensteadt
1Department of Pathology, Loyola University Medical Center, Maywood, IL 60153, USA.
Journal of Pharmaceutical Sciences
|June 1, 1995
Summary
The calibrant LHN-1 (F537) is too narrow for accurate low molecular weight heparin (LMWH) determination. Heparin mass calibrator (HMC) offers a wider range, providing more accurate molecular weight results for heparins.
Area of Science:
- Pharmaceutical Science
- Analytical Chemistry
- Biochemistry
Background:
- Accurate molecular weight (MW) determination is crucial for characterizing heparin and low molecular weight heparin (LMWH) therapeutics.
- High-performance size-exclusion chromatography (HPSEC) is a standard method for MW analysis of heparins.
- The suitability of calibrants directly impacts the accuracy of MW determination.
Purpose of the Study:
- To evaluate the performance of the calibrant LHN-1 (lot F537) for MW determination of heparins using HPSEC.
- To compare the accuracy of LHN-1 (F537) with a heparin mass calibrator (HMC) and a reference narrow-range-calibration method.
- To determine the optimal calibrant for accurate MW analysis across a range of heparin types.
Main Methods:
- High-performance size-exclusion chromatography (HPSEC) was employed for MW analysis.
- Molecular weight (MW), number-average MW, peak MW, and dispersity were calculated for various heparins.
- Calibrations were performed using LHN-1 (F537), heparin mass calibrator (HMC), and a reference narrow-range-calibration method.
Main Results:
- LHN-1 (F537) demonstrated a MW determination range too narrow for accurate analysis of low molecular weight heparins (LMWHs).
- Results obtained with HMC showed no significant difference from the reference method for substances up to 15.0 kDa.
- In contrast, LHN-1 (F537) yielded significantly different MW profiles compared to the reference method for samples above 7.5 kDa.
Conclusions:
- The calibrant LHN-1 (F537) is not suitable for accurate MW determination of all LMWHs due to its limited calibration range.
- Heparin mass calibrator (HMC) provides a broader and more accurate calibration range, suitable for both LMWHs and unfractionated heparins.
- HMC is a superior calibrant for HPSEC-based MW determination of heparins compared to LHN-1 (F537).