Prospective randomized comparison of biphasic waveform tilt using a unipolar defibrillation system
J E Poole1, G H Bardy, P J Kudenchuk
1Department of Medicine, University of Washington, Seattle, USA.
Pacing and Clinical Electrophysiology : PACE
|July 1, 1995
Summary
Lowering biphasic waveform tilt did not improve defibrillation efficacy in a unipolar transvenous system. This study found no significant difference in defibrillation threshold (DFT) between 50% and 65% tilt waveforms, suggesting current systems are adequate.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- A unipolar defibrillation system using a single right ventricular (RV) electrode and implantable cardioverter defibrillator (ICD) shell demonstrates comparable efficacy to epicardial lead systems.
- Optimizing pulse waveform shape may further enhance unipolar defibrillation system performance.
Purpose of the Study:
- To investigate if reducing biphasic waveform tilt improves defibrillation efficacy in a unipolar transvenous system.
- To compare the defibrillation threshold (DFT) using 50% and 65% tilt asymmetric biphasic waveforms.
Main Methods:
- Prospective, randomized comparison of 50% and 65% tilt biphasic waveforms in 15 cardiac arrest survivors.
- Utilized a unipolar system with an RV electrode (anode) and ICD shell (cathode).
- Measured defibrillation threshold (DFT) parameters including stored energy, delivered energy, voltage, current, resistance, and pulse width.
Main Results:
- The 65% tilt biphasic pulse yielded a stored energy DFT of 8.7 ± 5.7 J and delivered energy DFT of 7.6 ± 5.0 J.
- The 50% tilt biphasic pulse resulted in a stored energy DFT of 8.2 ± 5.4 J and delivered energy DFT of 6.1 ± 4.0 J.
- All patients in both groups achieved DFTs < 20 J, with no statistically significant difference between the two tilt angles (P = 0.69 and 0.17).
Conclusions:
- The unipolar single-lead transvenous defibrillation system achieves defibrillation at energy levels comparable to epicardial systems.
- Reducing biphasic waveform tilt to 50% does not enhance defibrillation efficacy compared to the standard 65% tilt.