Inconsistent response to rapid atrial rhythms in a DDD pacemaker with the fallback feature

N J Johnson1

  • 1Division of Cardiology, Case Western Reserve University School of Medicine, Cleveland, Ohio, USA.

Pacing and Clinical Electrophysiology : PACE
|July 1, 1995
PubMed
Summary

Dual chamber pacing can be problematic for atrial arrhythmias. The fallback feature

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Effects of ritanserin and 1-(2,5-dimethoxy-4-iodophenyl)-2-aminopropane (DOI) in the murine elevated plus-maze test of anxiety: an ethopharmacological study.

Journal of psychopharmacology (Oxford, England)·2012
Same author

Lung disease mortality in the United States: the National Longitudinal Mortality Study.

The international journal of tuberculosis and lung disease : the official journal of the International Union against Tuberculosis and Lung Disease·2009
Same author

Resumption of postpartum luteal function of primiparous, suckled beef cows exposed continuously to bull urine.

Journal of animal science·2006
Same author

Implantable cardioverter defibrillator (ICD) at end of life #112.

Journal of palliative medicine·2005
Same author

Cardiac pacemakers at end of life #111.

Journal of palliative medicine·2005
Same author

A validation of the 10-meter incremental shuttle walk test as a measure of aerobic power in cardiac and rheumatoid arthritis patients.

Archives of physical medicine and rehabilitation·2001

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Electrophysiology

Background:

  • Dual chamber pacemakers are used to manage bradyarrhythmias.
  • Paroxysmal atrial arrhythmias can pose challenges for pacemaker function.
  • The fallback feature aims to prevent pacemaker tracking of rapid atrial rhythms.

Observation:

  • A case study highlighted inconsistent initiation of the fallback feature during atrial flutter.
  • Inconsistent fallback can lead to inappropriate ventricular pacing.

Findings:

  • Fallback activation was consistently observed only when the atrial rate met a specific criterion.
  • The criterion for consistent fallback was 2xTARP (ms) <= URL (ms).

Implications:

Related Experiment Videos

  • This finding suggests a need for precise calibration of pacemaker settings.
  • Optimizing fallback parameters may improve device performance in patients with atrial arrhythmias.
  • Further research is warranted to refine pacemaker algorithms for complex arrhythmias.