Properties of hand-held spreaders after simulated use
1Section of Oral Biomaterials, Department of Basic Sciences, University of Texas-Houston Dental Branch 77225, USA.
Journal of Endodontics
|August 1, 1995
Summary
This study evaluated root canal spreader performance under simulated clinical use. While all tested spreaders met proposed standards, Endeco instruments showed greater susceptibility to permanent deformation after simulated use and sterilization.
Area of Science:
- Endodontics
- Dental Materials Science
- Biomaterials Engineering
Background:
- Root canal spreaders are critical endodontic instruments with varying dimensions.
- Existing standards (ANSI/ADA Specification no. 71) do not fully replicate clinical usage conditions for spreader testing.
- Understanding instrument behavior under simulated clinical stress is vital for ensuring endodontic treatment success and instrument longevity.
Purpose of the Study:
- To assess the permanent deformation of D11T root canal spreaders from different manufacturers.
- To evaluate the impact of simulated clinical usage (insertion cycles) and autoclaving on spreader integrity.
- To compare the performance of Endeco, Hu-Friedy, and Premier spreaders under various stress conditions.
Main Methods:
- Measured diameters (D3, D16) and taper of 40 D11T spreaders from three companies.
- Subjected spreaders to four conditions: control, autoclaving (10 cycles), simulated insertion (50 times, 200 mm/min into 30-degree curved blocks), and combined autoclaving/insertion.
- Quantified permanent angular deformation of the instrument tip using a bend test apparatus.
Main Results:
- All tested root canal spreaders successfully passed the bend test criteria under all experimental conditions.
- Endeco (EN) spreaders exhibited a higher susceptibility to permanent deformation compared to Hu-Friedy (HF) and Premier (PR) spreaders.
- No statistically significant differences in deformation were observed between the control group and the simulated clinical usage groups (B, C, D).
Conclusions:
- While all evaluated root canal spreaders meet proposed bend test specifications, material differences influence deformation resistance.
- Endeco spreaders may be less resilient to repeated stress and sterilization compared to competitors.
- Current testing protocols may not fully capture the long-term clinical performance and potential failure modes of root canal spreaders.