Evaluation of the pulse wave arrival time as a marker for blood pressure changes in critically ill infants and

C F Wippermann1, D Schranz, R G Huth

  • 1University Children's Hospital, Johannes Gutenberg University Mainz, Germany.

Journal of Clinical Monitoring
|September 1, 1995
PubMed

Insights

Pulse arrival time (PAT) does not reliably track blood pressure changes in critically ill infants and children. The correlation between 1/PAT and systolic blood pressure was insufficient for clinical use in this population.

Area of Science:

  • Biomedical Engineering
  • Pediatric Critical Care

Background:

  • Pulse arrival time (PAT) is the interval between the electrocardiogram (ECG) R wave and peripheral pulse wave arrival.
  • PAT has been explored as an indirect indicator of blood pressure fluctuations.

Purpose of the Study:

  • To assess the correlation between 1/PAT and blood pressure (systolic, diastolic, mean) and heart rate in critically ill infants and children.

Main Methods:

  • A laboratory device calculated PAT in real-time using ECG and pulse oximetry data.
  • Data from 15 critically ill patients (6 days to 16 years) were analyzed.

Main Results:

  • Mean correlation coefficients were highest for systolic blood pressure (r=0.73), followed by mean (r=0.68), diastolic (r=0.61), and heart rate (r=0.52).
  • Individual patient correlations for systolic blood pressure varied significantly (r>0.9 in 7, r<0.5 in 4).

Conclusions:

  • The correlation between 1/PAT and systolic blood pressure is not robust enough for monitoring blood pressure in critically ill pediatric patients.
  • Variations in the pre-ejection period, a component of PAT, may explain the weak correlation.