Related Experiment Videos
Improved method for measuring C1-r-C1-s-(C1 inh)2 complexes by an enzyme-linked immunosorbent assay
K P Mathews1, J H Herschbach, S L Chambers
1Department of Molecular and Experimental Medicine, Scripps Research Institute, La Jolla, California 92037, USA.
Abstract:
Measurement of C1-r-C1-s-(C1 inh)2 complexes in serum or plasma by enzyme-linked immunosorbent assay (ELISA) has been proposed as a relatively convenient and sensitive means for assessing C1 activation. However, interference by unactivated C1q (r-s)2 at low serum or plasma dilutions has resulted in estimates that vary widely with the degree of serum or plasma dilution. Precipitating the interfering C1q (r-s)2 with 6% polyethylene glycol has been proposed to resolve this problem, but here it is shown that this procedure also precipitates or coprecipitates some of the C1-r-C1-s-(C1 inh)2 complexes. Satisfactory results have been achieved without PEG precipitation by testing high plasma dilutions under conditions where there is a sufficient excess of anti-C1s coating the microtitration plate wells that removal of C1q (r-s)2 is not necessary. Optimizing conditions for quantitating these complexes at high dilution have been investigated. The mean normal EDTA plasma C1-r-C1-s-(C1 inh)2 complex measurement was 36.6 +/- 7.0 (S.D.) ELISA units with a 95% confidence interval of 19.5-47.6u. Besides providing a sensitive assay for C1 activation, measuring C1-r-C1-s-(C1 inh)2 complexes may help to clarify the pathophysiologic mechanisms resulting from C1 inh deficiency under various conditions.