Standard flow-time waveforms for testing of PEF meters

J L Hankinson1, R O Crapo

  • 1National Institute for Occupational Safety and Health, Division of Respiratory Disease Studies, Morgantown, WV 26505, USA.

American Journal of Respiratory and Critical Care Medicine
|August 1, 1995
PubMed
Summary

New flow-time waveforms improve peak expiratory flow (PEF) meter testing. Standard spirometer tests do not adequately assess PEF meters, leading to potential inaccuracies. These new waveforms offer a more precise evaluation method.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The need to change the method for defining mild airway obstruction.

The European respiratory journal·2011
Same author

Changes in the FEV₁/FVC ratio during childhood and adolescence: an intercontinental study.

The European respiratory journal·2010
Same author

Age- and size-related reference ranges: a case study of spirometry through childhood and adulthood.

Statistics in medicine·2008
Same author

Using the lower limit of normal for the FEV1/FVC ratio reduces the misclassification of airway obstruction.

Thorax·2008
Same author

Definition of COPD: based on evidence or opinion?

The European respiratory journal·2008
Same author

Interpretative strategies for lung function tests.

The European respiratory journal·2005

Area of Science:

  • Pulmonary Function Testing
  • Biomedical Engineering

Background:

  • The American Thoracic Society (ATS) recommends volume-time waveforms for spirometer testing.
  • Current ATS waveforms, though including peak expiratory flow (PEF) values, were not designed for PEF meter evaluation.
  • Existing methods using ATS waveform 24 lack the necessary flow-time waveform variety (rise times) for comprehensive PEF meter assessment.

Purpose of the Study:

  • To develop and validate a new set of flow-time waveforms specifically designed for evaluating peak expiratory flow (PEF) meters.
  • To establish a more accurate method for measuring PEF and related flow parameters like rise time and time to PEF.

Main Methods:

  • Developed a set of 26 specific flow-time waveforms for PEF meter evaluation.
  • Directly measured PEF from the new flow-time curves.

Related Experiment Videos

  • Compared PEF measurements derived directly from flow-time curves against PEF indirectly derived from ATS volume-time curves using the recommended algorithm.
  • Main Results:

    • A significant difference (up to 10.7%) was observed in PEF measurements between direct flow-time curve analysis and indirect volume-time curve analysis for waveforms with fast rise times.
    • Minimal differences in PEF derivation methods were noted for waveforms with slower rise times.
    • The 26 flow-time waveforms provide a standardized method for testing PEF meter software and hardware.

    Conclusions:

    • The developed set of 26 flow-time waveforms provides a superior method for testing peak expiratory flow (PEF) meters compared to existing ATS volume-time waveforms.
    • Accurate PEF measurement requires waveforms that capture a range of rise times, particularly for fast-rising flows.
    • These waveforms are crucial for the reliable testing of PEF meters and associated software algorithms.