The evaluation of the wave-form analysis capability of a new strain-gauge intracranial pressure MicroSensor
1Department of Clinical Neurosciences, Western General Hospital, Edinburgh, Scotland.
Abstract:
Both laboratory and clinical trials have indicated the reliability of a new catheter-tip, strain-gauge intracranial pressure transducer. We report on the results of bench tests comparing the wave-form analysis capabilities of the new Codman intracranial pressure MicroSensor and a similar strain-gauge transducer. Good agreement was found between the two sensors with no significant nonlinearity or hysteresis over a measured pressure range from 0 to 100 mg Hg. Under "fluid-filled" conditions, the MicroSensor showed good reproduction of an arterial pressure wave form; when spectral analysis was used to analyze the two sensor types and break them down into harmonic components, no significant differences could be found for any of the first 10 amplitude and phase harmonics. If proven to stay robust after long-term clinical use, this sensor may be a useful alternative to the existing transducer systems for routinely monitoring the intracranial pressure and its wave form.