Evaluation of a rapid readout biological indicator for 121 degrees C gravity and 132 degrees C vacuum-assisted steam
D Vesley1, M A Nellis, P B Allwood
1School of Public Health, University of Minnesota, Minneapolis 55455, USA.
Objective:
A biological indicator (BI), based on fluorescent detection of a spore-associated enzyme, has been developed. We evaluated the new BI against five conventional self-contained BIs in a 121 degrees C gravity-displacement sterilizer and, packaged in towel packs and disposable steam test packs, in a 132 degrees C vacuum-assisted sterilizer.
Design:
In the 121 degrees C gravity-displacement sterilizer, rapid readout BIs and conventional BIs were tested together in an otherwise empty chamber. Exposure times were 5, 10, and 15 minutes. In the 132 degrees C vacuum-assisted steam sterilizer, BIs were contained in towel packs or disposable steam test packs. Surgical packs were added to represent a full load. Exposure times were 0, 1, and 2 minutes.
Results:
In gravity-displacement 5-minute cycles, the rapid readout BIs yielded 100% positive results within 30 minutes. The conventional BIs ranged from 72% to 100% positive in 48 hours. At 10 minutes, the rapid readout BIs were 83% positive after 3 hours, whereas the conventional BIs ranged from 0% to 37% positive after 48 hours. All indicators were negative after 15-minute exposures. In vacuum-assisted cycles, all indicators were negative after 3 minutes. After 1-minute exposures, the rapid readout BIs were 40% to 60% positive, whereas the conventional BIs were 15% to 20% positive.
Conclusion:
The rapid readout BIs are a more sensitive indicator of marginal steam sterilization cycles than conventional self-contained BIs. They yield results within 3 hours, as compared to 24 to 48 hours for the conventional BIs.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Physical Methods for Controlling Microbial Growth: Temperature