Related Experiment Video
Updated: Aug 9, 2026

A Coupled Experiment-finite Element Modeling Methodology for Assessing High Strain Rate Mechanical Response of Soft Biomaterials
Published on: May 18, 2015
Axisymmetric finite element analysis of hip replacements possessing an elastomeric layer: the effects of clearance
1School of Engineering, University of Durham.
Abstract:
A finite element analysis has been performed for a hip replacement incorporating an elastomeric surface layer. The effects of assuming the elastomer as incompressible have been examined for imposed loads, by comparing the incompressible solution to that using accurately measured values for Poisson's ratio, and while deflections were found to be significantly different, the changes in maximum pressure in the contact, maximum shear stress, contact angle and angular position to maximum shear stress were all small (less than 10 per cent). Comparisons have also been made between the finite element results for ABAQUS and an asymptotic solution, and it was found to be important to use the numerical model particularly at small values of clearance.
Related Concept Videos
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...
Members Made of Elastoplastic Material
As the bending moment...
Plastic Deformations of Members with a Single Plane of Symmetry
Residual Stresses in Bending
Eccentric Axial Loading in a Plane of Symmetry