Polyethylene wear in total hip arthroplasty
A B Bankston1, H Cates, M A Ritter
1Orthopedic Clinic, Inc., Baton Rouge, LA, USA.
Clinical Orthopaedics and Related Research
|August 1, 1995
Summary
Polyethylene wear in total joint arthroplasty is influenced by fabrication method and metal backing, not femoral bearing material. Compression-molded all-polyethylene cups may reduce wear.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Biomedical Engineering
Background:
- Polyethylene wear is a primary cause of osteolysis and granuloma formation following total joint arthroplasty.
- Understanding factors influencing polyethylene wear is crucial for improving implant longevity.
Purpose of the Study:
- To evaluate polyethylene wear differences based on bearing surface materials (stainless steel, cobalt chrome, titanium alloy).
- To compare wear between cemented all-polyethylene molded cups and cemented metal-backed acetabular cups.
- To assess the impact of polyethylene fabrication (molded vs. machined) on wear rates.
Main Methods:
- Patient groups were matched for gender, age, weight, and follow-up duration.
- Linear wear rates were analyzed using the Livermore et al. technique.
- Evaluations included bearing surface material comparisons and acetabular component types (metal-backed vs. nonmetal-backed).
Main Results:
- Linear wear rates for bearing surfaces: stainless steel (0.06 mm/year), cobalt chrome (0.05 mm/year), titanium alloy (0.08 mm/year).
- Linear wear rates for acetabular components: metal-backed (0.011 mm/year), nonmetal-backed (0.08 mm/year).
- Femoral bearing material had minimal effect on polyethylene wear; fabrication method and metal backing were significant factors.
Conclusions:
- The metal composition of the femoral bearing surface has a negligible impact on linear polyethylene wear.
- Polyethylene fabrication method and the presence of metal backing significantly influence polyethylene wear rates.
- Compression-molded all-polyethylene acetabular components may be optimal for minimizing polyethylene wear in total hip arthroplasty.