Related Experiment Videos
Endoscopic electrode implantation. A new technique in an animal model
S G Osman1, G Polando, E B Marsolais
1Veterans Administration Medical Center, Cleveland, OH 44106-1702, USA.
Abstract:
The purpose of this study was 2-fold: (1) to describe a technique of creating an endoscopic interstitial working space; and (2) to describe the technique of implanting the modified double helix and the Huntington cuff electrode in the space created. Two cannulated balloons were inserted on either side of a standard arthroscope, which had been inserted to identify the target branch of the sciatic nerve. Traction was applied on each cannula after inflating the balloons with 3 to 5 cc of saline, to create a cylindrical operating space between them. After mobilizing a segment of the nerve, either of the 2 electrodes was implanted with the aid of special instruments. The results indicate that (a) it is possible to create interstitial operating spaces reliably and consistently; (b) satisfactory hemostasis can be achieved through the tamponade effect of the balloons and endoscopic cautery units; (c) excellent visualization is achievable; and (d) it is possible to implant electrodes in a reproducible manner. The interstitial spaces are portable and can be used for various procedures in most anatomic locations. Modification of the cuff electrode will be necessary to make it easier for endoscopic implantation.