Hydrophilic surface modification of metallic endoluminal stents
J M Seeger1, M D Ingegno, E Bigatan
1Department of Surgery, University of Florida, Gainesville, USA.
Purpose:
Stainless steel endovascular stents are inherently thrombogenic so that thrombus accumulates on these devices, leading to acute vessel occlusion. A potential solution to this problem is stent surface modification with hydrophilic polymers, which might limit platelet adhesion and reactivity.
Methods:
N-vinylpyrrolidone (NVP) and potassium sulfopropyl acrylate (KSPA) hydrophilic monomers were gamma graft polymerized onto 1 cm2 stainless steel slabs and 4 mm Palmaz stainless steel stents. Surface characteristics of modified and plain stainless steel stents were then investigated with contact angle and x-ray photoelectron spectroscopy measurements, and in vitro and in vivo platelet reactivity was assessed as 111Indium platelet accumulation expressed as counts/min/cm2.
Results:
Surface modification of stainless steel slabs and stents with both NVP and KSPA hydrophilic polymers significantly reduced in vitro platelet adhesion (plain = 2249 +/- 723 counts/min/cm2, NVP = 428 +/- 156 counts/min/cm2, KSPA = 958 +/- 223 counts/min/cm2) and in vivo platelet accumulation after 1 hour of blood flow exposure (plain = 1407 +/- 796 counts/min/cm2, NVP = 426 +/- 175 counts/min/cm2, KSPA = 399 +/- 124 counts/min/cm2. In addition, platelet accumulation on modified stents indexed to plain stents was lowest in KSPA-modified stents (NVP = 79.3% +/- 31.7% of plain, KSPA = 51.2% +/- 36.2% of plain). Surface analysis confirmed surface grafting with both monomers, and SEM documented smoothing of the irregular surfaces of the stainless steel stents after grafting.
Conclusion:
Hydrophilic polymer surface modification of stainless steel stents decreases initial stent surface platelet accumulation, which may decrease the risk of vessel thrombosis associated with the use of these devices.